405nm Raman Spectrograph
  • 300 – 3700 cm-1 spectrometer range
  • High NA, f/1.3
  • 10 μm slit :  8 cm¯¹
  • 25 μm slit : 12 cm¯¹
  • 50 μm slit : 20 cm¯¹
  • Free operating software & SDKs


Specification
OPTICAL
DETECTOR COOLING OPTIONS > Ambient Regulated TEC Cooled
Wavenumber Range (λex  = 405 nm) 300 - 3700 cm¯¹ 250 - 3000 cm¯¹
Resolution 10 μm slit 8 cm¯¹
25 μm slit 12 cm¯¹
50 μm slit 20 cm¯¹
f-number (f/#) 1.3
Connector (fiber coupled models only) SMA 905


 

DETECTOR  &  ELECTRONICS
DETECTOR COOLING OPTIONS > Ambient Regulated TEC Cooled
Hamamatsu Detector S10420-1006 CCD S16011-1006 CCD S10141-1007 CCD
Detector Temperature ambient 10°C -15°C
Detector Temperature Stability - ± 0.2°C ± 0.1°C

Active Pixels

1024 1024
Pixel Size 14 x 14 μm 12 x 12 μm
Integration Time 3 ms - 60 s 25 ms - 60 s
Maximum Sample Frequency 285 Hz
Communications USB 2.0 Type B connector


 

MECHANICAL & ENVIRONMENTAL
Fiber or Free Space Coupled
Size 16.5 x 16.2 x 6.7 cm
Weight 1.8 kg
Power Consumption <500 mA @ 12 V
Operating Temperature 0°C to 40°C, non-condensing


경기도 하남시 조정대로 150 하남지식산업센터(ITECO) 그린존 O213호
Tel : 02.486.7930 / Fax : 02.486.7931 / E-mail : kos@kosinc.co.kr
Copyright KOS. All rights reserved.