PRODUCTS
Solutions for Spectroscopy, TCSPC & Imaging
- Extended 270 – 3600 cm-1 spectrometer range
- High NA, f/1.3
- 10 µm slit : 7 cm-1
- 25 µm slit : 9 cm-1
- 50 µm slit : 13 cm-1
- Free operating software & SDKs
Catalog
Specification
OPTICAL | |||||
DETECTOR COOLING OPTIONS > | Regulated | TEC Cooled | |||
Wavenumber Range (λex = 785 nm , with longpass filter) | 100 - 3600 cm¯¹ | 200 - 3350 cm¯¹ | |||
Resolution (FWHM) | 10 μm slit | 6 cm¯¹ | 7 cm¯¹ | ||
25 μm slit | 7 cm¯¹ | 9 cm¯¹ | |||
50 μm slit | 12 cm¯¹ | 13 cm¯¹ | |||
f-number (f/#) | 1.3 | ||||
Connector (fiber coupled models only) | SMA 905 |
DETECTOR & ELECTRONICS | |||||
DETECTOR COOLING OPTIONS > | Regulated | TEC Cooled | |||
Hamamatsu Detector | S16010-1006 CCD | S16011-1006 CCD | |||
Detector Temperature | 10°C | -15°C | |||
Detector Temperature Stability | ± 0.2°C | ± 0.1°C | |||
Active Pixels | 2048 | 1024 | |||
Pixel Size | 14 x 14 μm | 24 x 24 μm | |||
Integration Time | 3 ms - 60 s | 8 ms - 60 s | |||
Maximum Sample Frequency | 285 Hz | 45 Hz | |||
Communications | USB 2.0 Type B connector |
MECHANICAL & ENVIRONMENTAL | |||
Fiber or Free Space Coupled | |||
Size | 24.0 x 13.5 x 7.1 cm | ||
Weight | 3.0 kg | ||
Power Consumption | <500 mA @ 12 V | ||
Operating Temperature | 0°C to 40°C, non-condensing |