High Resolution Spectrometer
200 - 1075nm

Resolutions better than 0.1nm

Specification
  Dynamic range:   2000:1 with 6 decades
  Optical resolution:   see model table – to 0.1 nm
  Detector type:   2048 pixel CCD/PDA
  Detector range:   190-1600nm
  Pixel size:   14 x 200um
  Diffraction Gratings:   Holographic & ruled
  Grating g/mm:   300, 600, 1200, 1800, 2400
  Spectrograph:   f/4, High Resolution SymX-Czerny–Turner
  Order sorting filters:   Integrated & High Pass
  Signal to noise:   1000:1
  Digitizer:   16-bit
  Dimensions:   Depends on specific model
  Power:   100mA @ 5 VDC
  Interface Options:   USB2 or Wifi (2), RS232 (3), SPI3, 4-20mA (3) Digital I/O (3) or ethernet (3)
  Data Transfer Speed:   30Hz or 1000Hz (1)
  Slit size options:   7, 14, 25, 50, 100, 200um or Interchangeable Upgrade
  Stray light:   <0.1% at 435nm; <0.05% at 600nm
  Fiber optic input:   SMA-905 0.22na single fiber
  Operating systems:   Windows and Linux (2), Andriod (2), iOS (2)
  Software included:   SpectraWiz Program, WinSDK (C,C#,VB, Delphi), Customizable LabVIEW, VBA for Excel (CRI & LED Report)


제품정보
High Resolution Spectrometer Models
  Model Wavelength
Range (nm)
Grating
g/mm
Grating
Range (nm)
Dispersion (nm/pixel) Slit-25
nm res.
Slit-14
nm res.
  HR-UV3   200-340   1800   140   0.068   0.3   0.2
  HR-UV3b   300-400   2400   100   0.048   0.21   0.14
  HR-UV4   200-300   2400   100   0.048   0.21   0.14
  HR-VIS   350-750   600   400   0.195   1.2   0.8
  HR-NIR   500-900   600   400   0.195   1.2   0.8
  HR-NIR2   900-1075   1200   175   0.085   0.26   0.17
  HR-NIR3   750-850   1800   100   0.049   0.15   0.1
  HR-NIR4   500-580   2400   80   0.039   0.12   0.08



LIBS Standard Resolution Spectrometer Models (LSR)

  Model Wavelength
Range (nm)

Grating
g/mm

Grating
Range (nm)
Slit-14
nm res.
  LSR-UV2   200-400   2400   200   0.2
  LSR-VIS4   400-600   2400   200   0.2
  LSR-VIS4b   600-800   1800   200   0.2
  LSR-VIS3b   800-1000   1800   200   0.2

*LIBS units come with Jack-In trigger, DLENS, & High Resolution Calibrations and when multiple units are ordered together have guaranteed start and stop wavelengths for continuous measurement


LIBS High Resolution Spectrometer Models (LHR)

  Model Wavelength
Range (nm)

Grating
g/mm

Grating
Range (nm)
Slit-14
nm res.
  LHR-UV3   200-300   2400   100   0.1
  LHR-UV3b   300-400   2400   100   0.1
  LHR-VIS3   400-500   1800   100   0.1
  LHR-VIS3b   500-600   1800   100   0.1
  LHR-VIS3c   600-700   1800   100   0.1
  LHR-NIR3   700-800   1800   100   0.1
  LHR-NIR2   800-975   1200   175   0.2

*LIBS units come with Jack-In trigger, DLENS, & High Resolution Calibrations and when multiple units are ordered together have guaranteed start and stop wavelengths for continuous measurement


NIR High Resolution Spectrometer Models

  Model Wavelength
Range (nm)
  InGaAs Pixel #   Slit-25nm res.
  DS- NIR3-HR   1530-1605   512   0.4
  DS- NIR3-HR   1500-1640   1024   0.4
  DS-NIR2   1250-1575   512   1.3
  DS- NIR2b   1150-1475   512   1.3
  NIRX   1500-2200   512   2.8
  NIRX   1500-2200   1024   1.4


경기도 하남시 조정대로 150 하남지식산업센터(ITECO) 그린존 O213호
Tel : 02.486.7930 / Fax : 02.486.7931 / E-mail : kos@kosinc.co.kr